PURPOSE
S: To develop a rapid, sensitive, qualitative ELISA-kit for serum adiponectin and examine correlation with adiponectin and cardiovascular risk factors.
METHODS
On the base of monoclonal antibodies against adiponectin, apply indirect ELISA to study the performance parameter of the kit. The correlation was examined between adiponectin and cardiovascular risk factors including waist circumference, body mass index, triglyceride, and HDL cholesterol.
RESULTS
The limited concentration of detection of the ELISA-kit was 1ug/ml. Linearity with R&D system and AdipoGen with this ELISA-kit was acceptable: the linear equation with R&D system was y=1.0116x + 0.4629 (R2=0.97) and linear equation with AdipoGen was y=0.9562x + 1.1961 (R2=0.93), respectively. The average recovery rate of the ELISA-kit ranged 92 to 104%. The correlation coefficient of waist circumference with adiponectin was -0.2276 (p<0.0001) among men and -0.2328 (p<0.0001) among women.
CONCLUSION
This ELISA-kit was quick, sensitive, and stable and can be used to determine adiponectin in serum.